Cath Lab Playbook
© Prabhu Sasankan, MD · Cardiology, Weill Cornell Medicine
Access

Cath Lab Playbook

already in placegoing in now (orange glow, arrow toward the patient)coming out now (faded, arrow toward you)

1. Before the case

RadiationLead apron, thyroid collar, dosimeter. Position the ceiling shield and table skirt before fluoro starts. Step back during cine runs when your hands are not needed.

2. Set up your table and manifold

T1

Lay out your table

A cardiac catheterization laboratory
Image did not load. View on Wikimedia Commons
Source: Wikimedia Commons.
A cath lab: C-arm over the table, monitors facing the operator.

Lay out the table so you can find things without looking:

  • A bowl of heparinized saline, for flushing and keeping wires and catheters wet
  • Wet Telfa or gauze, for wiping the wire every time it comes out
  • A wire bowl or hoop, for the coiled wire between uses
  • A waste bowl for aspirated blood, never the saline bowl
  • 10 and 20 mL syringes
T2

Learn the manifold

Stopcock pointsConnectsUsed for
RightCatheter to syringe; side line cut offInjecting, aspirating, passing through to the next port
DownSide line to the catheter; syringe cut offPort 1 at home: the transducer reads the catheter
UpSide line to the syringe; catheter cut offDrawing saline or contrast, pushing to waste, priming
LeftAll three openNot used
Example positions:Home: down, right, right (ports 1, 2, 3) + transducer stopcock right (closed to air). Pressure on screen, syringe cut off.
Draw saline / push out waste: down, up, right (ports 1, 2, 3).
Draw contrast: down, right, up (ports 1, 2, 3).
Flush transducer (once): up, right, right (ports 1, 2, 3) + transducer stopcock up (open to air).
Zero transducer (once): right, right, right (ports 1, 2, 3) + transducer stopcock up (open to air).
Inject: right, right, right. Straight from the syringe to the catheter.
T3

Prep the manifold

Out of the packet, every line is full of air. Before the manifold touches a catheter it must hold only saline and contrast, with no bubbles.

  1. Hang the saline bag, the contrast bottle and the waste bag. Squeeze and release the contrast controller until it holds contrast.
  2. Saline line: port 2 up, port 3 right. Pull: air comes first, then saline. Push it all into the waste bag. Repeat until the syringe fills with bubble-free saline.
  3. Contrast line: port 3 up, pull until contrast comes without air. Then port 3 right, port 2 up, and push it to waste.
  4. Flush transducer: draw bubble-free saline (port 2 up, pull, then port 2 right). Transducer stopcock up (open to air), port 1 up. Push gently until saline drips from the top with no bubbles. Transducer stopcock back to right.
  5. Flush through the front end of the manifold: right, right, right. Hold the syringe tip down and push saline until it drips out with no bubbles.
  6. Home: down, right, right. The manifold is primed and waits for the catheter.

Connecting the catheter and zeroing the transducer come once the catheter is in the aorta: section 5.

SimulatorPractice: prep the manifold

Starts fresh from the packet, at home position, every line full of air (white dashes). Work through the prep. The controls below the drawing are laid out like the manifold; with Show next step ticked (at the bottom), the control you need next glows.

Get readyPrep the manifold while the attending preps and drapes the access site, so it is ready before the first catheter needs it.

3. Access

Your access:

Pick one. From here through section 8 the guide shows only that access's steps and sizes. Until you pick, both are shown, tagged.

Sizes Needles are sized in gauge: the higher the number, the thinner the needle (21 gauge is about 0.8 mm; 18 gauge about 1.3 mm).
Wires are sized by diameter in inches: 0.014 (coronary), 0.018 and 0.021 (access), 0.035 and 0.038 (the workhorse J-wire). Each wire must fit through the needle or catheter it goes through.
Catheters and sheaths are sized in French: 1 French is 1/3 mm. A catheter's French is its outer diameter (6 French is 2 mm); a sheath's French is the largest catheter it takes.
A1

Sedation and pain relief

Moderate sedation, given in small doses and repeated as needed.

Watch the oxygen saturation, breathing and blood pressure on the monitor. The reversal drugs are naloxone (for fentanyl) and flumazenil (for midazolam).

CarefulSnoring, falling saturation or slow breathing means too much sedation. Say it out loud: jaw thrust, oxygen, and stop giving more.

Radial3A. Radial access

Arteries of the forearm
Image did not load. View on Wikimedia Commons
Arteries of the forearm (Gray's Anatomy, 1918). Source: Wikimedia Commons.
R1

Local anaesthetic

Lidocaine at the radial site.

Find the strongest pulse, 1 to 2 cm proximal to the radial styloid. With a small needle (25 gauge) raise a small wheal of 1% lidocaine in the skin right over it, 1 to 2 mL. Keep it small: a big bleb flattens the pulse you are about to aim for. Some attendings give it after a first look with ultrasound.

Get readyAccess needle or cannula out of the sheath kit, and the mini guidewire wet.
R2

Needle into the artery

21-gauge needle through the skin, bevel just inside the radial artery, pointing toward the elbow.

Arm on the board, wrist slightly extended, prepped and draped. Find the strongest pulse, 1 to 2 cm proximal to the radial styloid.

The needle, wire and sheath all come in the radial sheath kit (for example Terumo Glidesheath Slender). Your lab uses one of two techniques:

  • Anterior wall: bare 21-gauge metal needle, about 4 cm long, at 30 to 45 degrees. Stop as soon as bright red blood pulses out of the hub, and hold the needle still.
  • Through and through: a 20- or 22-gauge plastic cannula over a needle (like an IV catheter) goes through both walls. Take out the inner metal needle, then pull the plastic cannula back very slowly until blood pulses out. Hold it still.
Get readyHave the mini guidewire from the sheath kit (0.021 or 0.018 inch, about 45 cm) out of its packet and wet before the needle goes in.
R3

Mini guidewire through the needle

Mini guidewire, 0.021 inch (0.018 inch in some kits), threaded through the needle into the artery.

The mini guidewire is short (about 45 cm) and thin, with a soft straight tip; it only holds the track for the sheath, and is too thin to carry a catheter. Thread it through the needle or cannula. It should slide with almost no resistance. Advance it until only enough is left outside to hold it and load the sheath.

CarefulResistance means a side branch, a loop or spasm. Stop. Pull back a little, check for blood return, or ask for help. Forcing it causes perforation and forearm hematoma.
R4

Needle out, wire stays

The needle slides back toward you off the wire. The wire stays in the artery.

Hold the wire at the skin with one hand and slide the needle back off the wire's back end with the other. If your lab does it, make a small skin nick beside the wire.

Get ready6 French radial sheath, 10 or 16 cm long (for example Glidesheath Slender), flushed, with its dilator locked inside it.
R5

Sheath and dilator over the wire

6 French sheath, with its dilator locked inside, sliding over the 0.021 inch wire into the artery.

Thread the wire's back end into the dilator tip and slide the sheath down the wire to the skin. Before the sheath enters the skin, the wire's back end must be coming out of the back of the dilator, with you holding it. Then advance sheath and dilator together, all the way to the hub.

A sheath's French size is the largest catheter it takes: a 6 French sheath takes a 6 French catheter. "Slender" sheaths have a thinner wall, so a 6 French Slender sheath is about as wide on the outside as a standard 5 French sheath, which matters in a small radial artery.

R6

Dilator and wire out, sheath stays

Dilator and wire come out together. Only the sheath stays.

Unlock the dilator from the sheath and pull the dilator and wire out together.

Get readyA fresh 10 mL syringe of heparinized saline for the sidearm, and the radial cocktail drawn up and labelled.

Femoral3B. Femoral access

F1

Find the target on fluoro

Inferior epigastric artery(lowest point marks the ligament)External iliacTarget: common femoral arteryover the middle third of the femoral headClamp on the skin marks the mid-headFemoral headBifurcation (below the head)Superficial femoralProfundaToo high: retroperitoneal bleedToo low: pseudoaneurysm, AV fistulaR
Right groin on AP fluoro: the target is the common femoral artery over the middle third of the femoral head.

Feel the pulse, then mark the middle of the femoral head on fluoro with a clamp on the skin. Find the common femoral artery with ultrasound if your lab uses it (most do). The skin crease is not a reliable landmark.

"Micropuncture":The end goal is to get a 0.035 inch wire in the artery, and then get the main sheath in over that.

Micropuncture splits it into two stages:
1. Find the artery and make a small hole.
Using a 21-gauge needle + 0.018 inch wire.
If: the wire runs the wrong way on fluoro, or the femoral angiogram shows the stick is above the inferior epigastric artery or below the bifurcation, pull everything out and hold pressure (a 21-gauge hole usually seals easily).

2. Upsize from there.
Insert the micropuncture introducer (with its 3 Fr dilator inside) over the 0.018 wire.
Then remove the dilator and the 0.018 wire.
Then insert the 0.035 inch wire.
Then insert the femoral sheath (with its own dilator) over that.
F2

Local anaesthetic

Lidocaine at the femoral site.

Over the target, raise a skin wheal of 1% lidocaine, then advance the needle along the track you will use and infiltrate down toward the artery, 10 to 15 mL in total. Pull back on the syringe before each injection, so none goes into a vessel. Some labs nick the skin with a scalpel now; others after the wire is in.

Get ready21-gauge micropuncture needle and the 0.018 inch wire out and wet.
F3

Micropuncture needle in

21-gauge micropuncture needle, 7 cm long, into the common femoral artery, pointing toward the head.

The micropuncture needle is often echogenic (textured to show up on ultrasound). Advance at about 45 degrees toward the target until blood pulses out of the hub. Stop and hold the needle still.

Get ready0.018 inch micropuncture wire (about 40 cm) out and wet.
F4

0.018 inch wire through the needle

0.018 inch micropuncture wire, about 40 cm, through the needle.

Thread the wire through the needle, then check its course on fluoro: it should run up along the spine, not off laterally or into a small branch.

F5

Needle out

Needle slides back off the wire; the 0.018 wire stays.

Hold the wire at the skin and slide the needle off its back end. Compress lightly.

Get readyMicropuncture introducer: a 4 or 5 French introducer with a 3 French dilator locked inside it.
F6

Introducer and dilator over the wire

4 or 5 French micropuncture introducer, dilator locked inside, going over the 0.018 inch wire into the artery.

Thread the wire's back end into the dilator tip , hold the wire where it comes out the back, and advance the dilator into the artery to its hub.

CarefulThe 0.018 inch wire is thin and floppy, and it is in the artery. Keep it still at the skin while you load: do not let it slide in or pull out.
F7

Dilator and 0.018 wire out

3 French inner dilator and 0.018 inch wire come out together; the 4 or 5 French introducer stays.

Unlock the inner dilator and pull it out with the 0.018 wire. Many attendings shoot a femoral angiogram now, with a small contrast injection through the introducer in an ipsilateral oblique view; others do it later through the sheath sidearm.

Get readyThe short wire from the sheath pack: 0.035 or 0.038 inch, about 45 cm.
F8

0.035 inch wire through the introducer

Sheath-pack wire, 0.035 inch, threaded through the introducer.

Check on fluoro that it runs up along the spine.

F9

Introducer out

Introducer slides back off the 0.035 wire; the wire stays.

Hold the wire and slide the introducer off. Compress.

Get ready6 French femoral sheath, 10 or 11 cm (for example Terumo Pinnacle or Cordis Avanti), flushed, dilator locked inside. A 23 cm or longer sheath if the iliacs are tortuous.
F10

Sheath and dilator over the wire

6 French sheath with its dilator going over the 0.035 inch wire into the artery.

Thread the wire's back end into the dilator tip , hold the wire where it comes out the back, and advance sheath and dilator together to the hub.

F11

Dilator and wire out, sheath stays

Dilator and 0.035 wire come out together; only the sheath stays.

Then flush the sidearm: step S1 below. Keep a hand on the sheath until it is secured.

Get readyThe 0.035 inch, 150 cm J-wire wiped wet, and the first diagnostic catheter flushed: a JL4, 5 or 6 French, 100 cm.

Both: flush the sheath

S1

Flush the sheath

The sheath's sidearm ends in a white stopcock. On this kind, the long lever (often marked OFF) points at the port that is closed; the other two ports are connected. Its three ports: the tubing to the sheath (the patient), the syringe port opposite it, and an uncapped side port at 90 degrees to both. At rest the OFF lever points at the patient. Radial and femoral sheaths are flushed the same way.

  1. With OFF at the patient, attach a fresh 10 mL syringe of heparinized saline to the syringe port.
  2. Turn 180 degrees: OFF at the syringe. The sheath now connects to the side port. Let blood run back out of the side port for a moment. That clears air and stagnant blood without it going near the syringe.
  3. Turn 90 degrees: OFF at the side port. The syringe now connects to the sheath. Draw back a little, until you see some blood in the syringe: the line is continuous and free of air.
  4. Flush forward into the sheath.
  5. Turn 90 degrees back: OFF at the patient. Keep pushing: saline now goes out of the side port and clears the blood from it. Leave the last mL in the syringe.

Practice: flush the sheath

The artery is on the left, the stopcock and syringe toward you on the right. Work through the flush. Do it with the controls below the drawing; with Show next step ticked (at the bottom), the control you need next glows.

R7

Radial cocktail

Three drugs into the sidearm, one syringe.

Give it through the sidearm: with OFF at the side port, the syringe connects to the sheath; turn OFF back to the patient afterwards. A common version: verapamil 2.5 mg, nitroglycerin 100 to 200 mcg, and heparin (often 5,000 units or about 50 units/kg for a diagnostic case). Draw blood into the syringe to dilute it, inject slowly, and warn the patient that it will burn. Secure the sheath the way your lab does.

Get readyThe 0.035 inch, 150 cm J-wire wiped wet, and the first diagnostic catheter flushed: a JL3.5, 5 or 6 French, 100 cm (or the Tiger, if your attending starts with a universal catheter).
Continue: section 4 ↓

4. The diagnostic run: plan, and catheter up to the aorta

Your access:

Pick one. Sections 4 to 8 then show only that access's sizes and steps. Until you pick, both are shown and tagged.

A standard diagnostic left heart cath runs in this order. Each catheter goes up to the ascending aorta the same way (steps C1 to C6), is connected and zeroed once (section 5), and is exchanged over the J-wire (section 7).

StepWhat you useWhere
SheathRadial6 French Glidesheath Slender, 10 or 16 cm. Heparin went in with the radial cocktail.Femoral6 French sheath, 10 or 11 cm. Usually no heparin for a diagnostic femoral case; the attending decides.Radial3AFemoral3B
Wire0.035 inch J-wire, 150 cm, for every catheter.Radial Some attendings exchange over a 260 cm exchange-length wire from the radial (section 7).4
1. Left coronaryRadialJL3.5, 5 or 6 French, 100 cm (most attendings). One attending starts with a universal catheter (Tiger) that does both coronaries.FemoralJL4, 5 or 6 French, 100 cm.6
2. Right coronaryJR4, 5 or 6 French, 100 cm.Radial Skipped if the Tiger engaged both.7
3. LV and aortaAngled pigtail, 5 or 6 French, 110 cm. LVEDP in most cases. LV gram and aortogram only when indicated.8
Out and closureRadialCompression band (TR Band).FemoralManual compression or a closure device.9

A catheter's French size must be no larger than the sheath's: 5 or 6 French catheters go through a 6 French sheath. Ask the attending which size before you flush.

C1

Load the J-wire into the catheter

J tip1Right hand: middle, ring and little fingers grip the shaft.Thumb and index pinch the wire separately,with the J tip just beyond them.pull apartJ now straight2Pinch, and pull the two grips apart.The stretch between them straightensthe J tip beyond your thumb and index.catheterhubfeed in3Keep it straight and feed the tipinto the catheter hub.catheter tipwire tip: stop herewire inside catheter4Push the wire through until its tipis just inside the catheter tip.
On the table, before anything goes into the patient.

First catheter: RadialJL3.5 (or the Tiger, if your attending starts with a universal catheter), 5 or 6 French, 100 cm.FemoralJL4, 5 or 6 French, 100 cm.

Flush the catheter until saline drips from its tip. Wipe the 0.035 inch, 150 cm J-wire wet. Straighten the J with one hand: grip the shaft with your middle, ring and little fingers, and pinch the wire with your thumb and index just behind the J. Pull the two grips apart; the stretch straightens the J tip beyond your thumb and index. Keep it stretched and feed the straight tip into the catheter hub. Push the wire through. Stop with the wire tip just inside the catheter tip.

C2

Catheter and wire into the sheath

skin (patient)← toward the heartarterysheath valvesidearm: leave it as it isgripyour hand: 3 to 5 cm from the valvepush in a few cm, short strokesinside: wire tip stays just inside the cathetertable (you) →
Catheter, with the J-wire inside it, going through the sheath valve together.

Hold the catheter 3 to 5 cm from the sheath entrance, and push catheter and wire in together, in short strokes, until both are a few centimeters inside the sheath. The wire tip stays just inside the catheter tip while they enter the sheath; the wire only goes ahead in the next step.Radial The arm is on a board: keep the rest of the catheter flat on the drape, not hanging off the edge, so it does not pull back out.

C3

Wire leads up to the ascending aorta

Wire often drops here.Pull back, deep breath,try again.catheter waits hereFrom the right radial sheathcatheter waits hereFrom the right femoral sheathJ-wirecatheterAscending aortaArchDescending aortaAortic valveRight subclavianBrachiocephalicRight iliacR
AP view. The J-wire leads, the catheter stays behind it. RadialFrom the right radial: up the arm, subclavian and brachiocephalic, then down into the ascending aorta.FemoralFrom the right femoral: up the iliac and the aorta, around the arch, into the ascending aorta.

RadialHold the catheter still and advance the wire alone under fluoro. Watch the tip at the elbow and at the shoulder. At the shoulder it should turn down into the ascending aorta. It often drops into the descending aorta instead: pull it back into the subclavian, ask the patient to take a deep breath and hold it (that straightens the angle), and advance again. The attending may turn the catheter to point the wire.

FemoralHold the catheter still and advance the wire alone under fluoro, up the iliac and the abdominal aorta. Watch the tip, not the shaft. Over the arch it goes into the ascending aorta toward the valve.

CarefulIf the wire stops or bends, stop pushing. RadialIn the arm it may be in a small branch, a radial loop or spasm. Pull back a little and redirect; the attending may inject a few mL of contrast through the catheter to see the arm vessels.FemoralIt may be in a small branch or a tortuous iliac. Pull back a little and redirect, under fluoro.
C4

Catheter follows the wire

From the right radial sheathFrom the right femoral sheathcatheter tip just behind the JJ-wirecatheterAscending aortaArchDescending aortaAortic valveRight subclavianBrachiocephalicRight iliacR
AP view. Wire held still; the catheter slides up over it until its tip is just behind the J.

The catheter tip never passes the wire tip.Radial In a tortuous subclavian the catheter may not follow around the turn. Another deep breath often helps; if not, the attending may switch to a hydrophilic wire (Glidewire, 0.035 inch, 150 cm).Femoral Bring the catheter around the arch into the ascending aorta.

Get readyWet gauze in your hand. "Wire out" comes next.
C5

Wire out

wire coming outFrom the right radial sheathwire coming outFrom the right femoral sheathtip takes its curveJ-wirecatheterAscending aortaArchDescending aortaAortic valveRight subclavianBrachiocephalicRight iliacR
AP view. Catheter held still in the ascending aorta while the wire comes out; without the wire inside it, the catheter tip takes its curve.

Hold the catheter hub still and pull the wire out through the wet gauze, coiling it as it comes.

Get readyAn empty 10 mL syringe for the catheter hub.
C6

Aspirate the catheter and connect the manifold

catheter (to patient)pullempty 10 mL syringe: blood, no bubbles1Empty syringe straight on the hub. Pull untilblood comes with no bubbles. Take it offand discard the blood (waste bowl).hub drips bloodred rotating adaptermanifold (saline dripping)wet to wet2Let the hub drip while the manifolddrips saline, then connect them,drop to drop, so no air gets in.
At the catheter hub, on the table.

The catheter hub has no stopcock. Put an empty syringe straight onto the hub, pull back until blood comes with no bubbles, take it off, and discard the blood. Let the hub drip while you bring over the manifold's red rotating adapter, and connect wet to wet. First catheter only: connect and zero next (section 5).

5. Connect catheter to manifold and zero

Z1

Zeroing

With the catheter connected, the manifold still has to be zeroed before the pressure means anything. On your manifold the transducer sits on port 1 with its own small stopcock on top.

  1. Check the transducer cable has been handed off the table and plugged into the monitor.
  2. Position the transducer at mid-chest height (the phlebostatic axis).
  3. Port 1 right: the transducer is cut off from the catheter.
  4. Transducer stopcock up: the transducer is open to air.
  5. Press ZERO on the monitor. The reading goes to 0.
  6. Transducer stopcock back to right (closed to air).
  7. Home: down, right, right. The aortic pressure appears.

Re-zero if the table height changes.

SimulatorPractice: connect, zero, go home

Starts with a primed manifold and the catheter waiting in the aorta. Now connect it, zero, and go home. Do it with the controls below the drawing; with Show next step ticked (at the bottom), the control you need next glows.

Get readyEyes on the pressure tracing from now on.

6. Left coronary: engage, check pressure, inject

E1

Engage the left coronary and check the pressure

124/76AONormalsharp upstroke, dicrotic notch96/84AODampedlow, rounded, no notch118/12AOVentricularizeddiastole falls toward zero
After engagement: a normal aortic trace (left); damped and ventricularized traces (right) mean stop.

RadialJL3.5 from the right radial. The attending advances it toward the left sinus in a shallow LAO view. It often needs a gentle push with a small turn, and a deep breath, to drop into the left main. Coming down from the right subclavian, a catheter behaves about half a size larger than from the femoral; that is why radial starts with JL3.5 where femoral starts with JL4.

RadialTiger (universal), if that is your attending's choice. Advanced into the left sinus, it engages the left main with little or no rotation. It stays in for the right coronary too (section 7).

FemoralJL4 from the femoral. The attending advances it slowly in a shallow LAO view. With the wire out, the JL4 usually drops into the left main by itself as it is advanced. If its tip floats above the ostium and will not reach, the next size up (JL5); if it folds or points down, the next size down (JL3.5).

Compare the tracing with the aortic pressure you recorded. If it damps or ventricularizes, say so out loud and do not inject: the tip is against the wall or there is ostial disease, and the attending will reposition. If it is normal, a small test injection confirms position, then the views.

E2

Load, inject, go home

Tip down, toward the manifoldBubbles rise to the plungerand stay thereNever empty the syringe;leave the last few mL
Syringe tip down while injecting: any air rises to the plunger and stays behind the contrast.
  1. Load: from home, turn port 3 up and pull back slowly. Port 1 stays down, so you keep watching pressure.
  2. Tip the syringe down and tap so bubbles rise to the plunger. If there is a lot of air: port 3 right, port 2 up, syringe tip up, push the air to the waste bag, port 2 right, reload.
  3. Inject: right, right, right. The pressure trace goes flat while you inject; that is expected.
  4. Inject with the tip down: a smooth push, commonly 6 to 10 mL for the left coronary. Never empty the syringe; leave the last few mL.
  5. Home: down, right, right. Check the tracing before the next injection.
SimulatorPractice: one injection

Starts at home, catheter engaged and zeroed. Now load contrast, inject, and go back home. Do it with the controls below the drawing; with Show next step ticked (at the bottom), the control you need next glows.

Get readyKnow the next view before the injection ends; the attending will move the C-arm right away.
E3

Aspirate through the manifold

When blood has sat still in the catheter, or the trace looks damped from a clot or bubble, aspirate through the manifold before the next injection: right, right, right, pull until blood comes with no bubbles, then port 2 up and push it into the waste bag, then home.

SimulatorPractice: aspirate and discard

Starts at home with the catheter connected. Now aspirate, send the blood to waste, and go home. Do it with the controls below the drawing; with Show next step ticked (at the bottom), the control you need next glows.

E4

Left coronary views

Usually five or six views. Common order and what each shows:

ViewBest for
RAO caudalLeft main, proximal LAD, circumflex and OMs
AP caudalLeft main, proximal circumflex
LAO caudal ("spider")Left main bifurcation, proximal LAD and circumflex
AP cranialMid and distal LAD, diagonals
RAO cranialMid and distal LAD, diagonal origins
LAO cranialLAD and diagonals separated, septals
Diagram of the coronary arteries
Image did not load. View on Wikimedia Commons
Coronary anatomy. Source: Wikimedia Commons.
Coronary angiogram
Image did not load. View on Wikimedia Commons
Coronary angiogram with a stenosis. Source: Wikimedia Commons.
Get readyBefore the last left view: J-wire wiped and ready to straighten, and the JR4 (5 or 6 French, 100 cm) flushed. If your attending is using the Tiger, no new catheter: go to X5.
Get readyBefore the last left view: J-wire wiped and ready to straighten, and the JR4 (5 or 6 French, 100 cm) flushed.

7. Exchange and right coronary

RadialTiger: no exchange. After the left views the attending pulls back slightly out of the left main and rotates clockwise; the tip drops into the right coronary. Go to X5. If it will not engage the right, exchange to a JR4 as below.

X1

Wire out past the catheter tip

J-wire through the catheter and out its tip into the ascending aorta.

The attending disengages the catheter. Disconnect it from the manifold. Reinserting the J-wire: re-wet the J-wire, straighten the tip (pinch and pull), feed it into the catheter end, and advance under fluoro until the J tip emerges from the catheter tip in the ascending aorta.

RadialIf your attending uses a 260 cm exchange-length wire for radial exchanges, this is where it goes in. It is long enough to hold at its back end while the whole catheter comes off, and it stays in the ascending aorta, so the next catheter does not have to find its way around the subclavian again.

X2

Pin the wire, pull the catheter

Wire pinned at its back end; catheter pulled back toward you.

Hold the wire still at the catheter end and pull the catheter back.

X3

Grab the wire at the sheath

Catheter tip has cleared the sheath valve; your other hand takes the wire right at the sheath.

As you pull the catheter back, once the wire is exposed between the catheter tip and the sheath, grab the wire there.

CarefulWith a 150 cm wire in a 100 cm catheter this moment comes soon after you start pulling. If nobody holds the wire at the sheath here, it comes out with the catheter.
X4

Old catheter off, JR4 on

Old catheter slides off the wire's back end; the wire is held at the sheath.

Wipe the wire as the old catheter comes off. Thread the wire's back end into the JR4's tip , and slide the catheter on. Keep going until the wire emerges from the catheter end, then hold the wire there like a rail; advance the JR4 into the sheath, and then let the wire begin leading as you advance back up to the aorta. Then wire out , aspirate the hub, connect the manifold, and go home. No need to re-zero.

X5

Engage the right coronary and inject

Bring the catheter to just above the valve in the LAO view, then pull back slightly while rotating it clockwise, about 90 to 180 degrees. The tip turns toward the right sinus and drops into the RCA ostium.

RadialFrom the radial, torque takes longer to reach the tip. The attending turns slowly and lets the tip catch up; turned too fast, it whips past the ostium.

Check the pressure before each injection, exactly as for the left. Injections are smaller: commonly 4 to 6 mL.

ViewBest for
LAOProximal and mid RCA
RAOMid RCA, PDA length
AP cranial (or LAO cranial)Crux, PDA and posterolateral branches
CarefulDamping is common on the right: the tip can sit deep, or in the conus branch. If the trace damps or ventricularizes, say so and do not inject. A forceful injection into a damped catheter can dissect the artery or cause VF.
Get readyBefore the last right view: the pigtail (5 or 6 French, 110 cm) flushed and the J-wire ready. If an LV gram or aortogram is planned, ask the tech for the injector settings.

8. Pigtail: LVEDP, LV gram, aortogram

When to do which

LVEDP. Measured in most diagnostic cases. It gives the filling pressure before more contrast or fluid goes in. Skipped when the valve should not be crossed (see below).

LV gram (always with a pigtail). Do it when:

  • LV function is unknown: no recent echo, or the echo images are poor
  • Wall motion or mitral regurgitation needs assessing and the echo did not answer it
  • The attending wants to match wall motion to the coronary findings

Aortogram (pigtail in the aortic root). Do it when:

  • Aortic regurgitation needs grading and the echo did not answer it
  • Bypass grafts or an anomalous coronary will not engage, to show where they arise
  • The aortic root or ascending aorta needs assessing (dilation) and no CT or MRI already shows it

Skip the LV gram, or ask first, when:

  • A recent echo already answers the question
  • Contrast is limited by kidney function: an LV gram uses about 30 to 40 mL
  • LVEDP is very high, or the patient is in decompensated heart failure: the contrast load can tip them into pulmonary edema
  • There is a known LV thrombus

Do not cross the valve with a mechanical aortic valve. In severe aortic stenosis the attending crosses only if the gradient is needed, with a straight wire. With left bundle branch block, the catheter can bump the right bundle and cause complete heart block: know where temporary pacing is.

L1

Exchange to the pigtail

Exchange as in section 7 to an angled pigtail, 5 or 6 French, 110 cm. To load it, straighten the loop a little with your fingers so the end hole points at the wire, and load it with the fingertip technique . Every LV gram and aortogram is done with a pigtail: the loop keeps the tip off the ventricular wall, and the side holes spread the contrast. Leave the tip in the ascending aorta, just above the aortic valve. Wire out , aspirate, connect, home.

RadialThe pigtail comes down the ascending aorta toward the valve the same way. In a tall patient, or with a long arm route, 110 cm may not reach; ask for a longer one.

L2

Clear contrast from the line

Contrast in the line damps the pressure. Port 2 up, pull 5 to 10 mL of saline, port 2 right, then right, right, right and push it through, then home.

L3

Cross the valve, record LVEDP

The attending crosses the aortic valve into the LV with the pigtail. Watch the monitor for ectopy. Record LV pressure. LVEDP is read at end diastole: just after the small a wave, right before the steep upstroke of systole.

CarefulFrequent PVCs or a run of VT means the pigtail is touching the wall. Say it out loud; the attending pulls back slightly.
L4

LV gram (only if indicated)

  1. Disconnect the pigtail from the manifold. Connect the power injector line wet to wet, with no air in the connection.
  2. View: usually RAO 30 degrees (some add LAO 45 to 60 for the septum and lateral wall).
  3. The tech sets the injector. Common settings are about 30 to 36 mL at 10 to 15 mL/s; your lab's defaults rule.
  4. Watch for ectopy during the injection: PVC beats make that part of the gram hard to read.
  5. Reconnect the manifold, home. The attending may want a repeat LVEDP.
L5

Pullback

The pigtail is pulled back from the LV across the valve into the aorta while the pressure is recorded. A drop in systolic pressure as it crosses the valve (a gradient) suggests aortic stenosis. A drop while still inside the LV suggests obstruction below the valve.

L6

Aortogram (only if indicated)

  1. Pigtail tip in the aortic root, 2 to 3 cm above the valve (above the sinuses), so contrast fills the root without going through the valve.
  2. Power injector connected wet to wet. Common settings are about 40 mL at 20 mL/s; your lab's defaults rule.
  3. The attending picks the view, often LAO, to open out the root and show any regurgitant jet into the LV.
Get readyNext: catheter out and closure (section 9). RadialCompression band ready.FemoralClosure device or compression ready, and the ACT result back.

9. Catheter out and closure

D1

Catheter out

Ask whether the attending wants the wire back in before the catheter comes out. Many do for curved catheters coming down the arm, so the tip does not catch the vessel wall. If yes: wire in until its tip reaches the catheter tip, then pull catheter and wire out together, under fluoro if there is resistance. The sheath stays until closure.

Get readyCompression band open (radial), or the closure device open and the ACT result back (femoral).
D2

Radial: compression band

  1. If your band's instructions say so, pull the sheath back 2 to 3 cm.
  2. Place the band so its marker sits over the arterial puncture, a few millimetres proximal to the skin entry. Close the strap.
  3. Inflate the band (for example Terumo TR Band) with its own syringe, commonly 15 to 18 mL of air, per your unit's protocol, while you slowly pull the sheath out.
  4. Let air out slowly until a little blood appears at the site, then put 1 to 2 mL back until it stops. That is patent hemostasis: no bleeding, radial artery still open.
  5. Check: pulse oximeter on the thumb, compress the ulnar artery; a waveform means the radial is open.
D3

Femoral: compression or closure device

Manual compression: the ACT usually needs to be below a set value (often 150 to 180 seconds). Find the pulse 1 to 2 cm above the skin entry, pull the sheath in one smooth motion, and press firmly with two or three fingers over the arterial puncture. Hold firm pressure, then ease gradually, usually 15 to 20 minutes in total. Check the site and the distal pulses.

Closure devices such as Angio-Seal (6 or 8 French: an anchor inside the artery and a collagen plug outside) or Perclose ProGlide (a suture across the puncture) need a femoral angiogram showing a suitable puncture site; follow the device's steps and the attending.

10. Problems to spot early

For each one: notice it, and say it out loud.

What you seeWhat it might be
Wire or catheter won't advance up the arm; patient in painRadial spasm, loop, or anomaly
Pressure damps or ventricularizes after engagementCatheter against the wall, or ostial stenosis
Contrast stays in the vessel wall after injectionDissection
Slow or absent flow after stentingNo-reflow, dissection, or thrombus
Bradycardia, hypotension, nausea, sweatingVasovagal reaction
Contrast outside the vesselPerforation
Hives, wheeze, hypotension after contrastContrast reaction
Hypotension with back or flank pain after femoral accessRetroperitoneal bleed
Sudden ST elevation after an injection, bubbles seenAir embolism

Loading onto the wire: the fingertip technique

wire endfingertip1Lay the wire's end on yourfingertip. Nothing hangsover the edge.device tip2Set the device tip down onthe same fingertip, linedup with the wire end.3Nudge the tip onto the wire.The wire slides intothe hole.

When: whenever the back end of a wire has to go into the tip of a device. The sheath dilator over the access wire, the micropuncture introducer over the 0.018 inch wire, each new catheter (and the pigtail) over the J-wire, and every balloon or stent over the 0.014 inch coronary wire.

  1. Wipe the wire end wet.
  2. Lay the wire's end flat on the pad of your index fingertip. No overhang: the end sits on the finger, not past its edge.
  3. With your other hand, set the device tip down on the same fingertip, pointed straight at the wire end.
  4. Nudge the device tip onto the wire end. The finger holds both at the same level and in line, so the wire finds the hole.
  5. Once a few millimetres are in, slide the device on along the wire.

Why it works: a wire end held in the air and a device tip held in the air each wobble on their own, and the holes are tiny (a balloon tip fits a 0.014 inch wire). Resting both on one fingertip puts them in the same plane, so they cannot drift apart.

Wipe and wind: taking the wire out

catheter (to patient)hubwet gauzewire comes outwire bowl (saline)4Aspirate thecatheter, thenmanifold1Gauze around wireat the hub; wipeas it comes out2Coil 15 to 20 cmloops in yourother hand3Whole coil intothe bowl
  1. While the catheter travels up, pick up a wet Telfa or gauze square.
  2. When the catheter tip is in the ascending aorta, the attending (or you) says "wire out."
  3. Hold the catheter hub still with one hand so the catheter does not move.
  4. With the other hand, wrap the gauze around the wire where it leaves the hub and pull the wire out through it in long, smooth strokes. The gauze wipes the blood off.
  5. Coil the wire into 15 to 20 cm loops in your free hand as it comes.
  6. When the J tip leaves the hub, put the whole coil in the wire bowl, with nothing hanging off the table.
  7. Immediately aspirate the catheter hub and connect the manifold.

Why wipe: blood on a wire clots, the wire becomes sticky and hard to handle, and the clot can be carried back in on the next pass.